SRP-001 SmartReader Plus 1 Temperature Data Logger

ACR 2 Channel Temperature Data Logger with 1 External Input
SKU
01-0008
$539.00
  • Internal Ambient Temperature Channel and External Thermistor Input
  • -40°C to 70°C (-40°F to 158°F) Temperature Measurement Range with 0.2°C Accuracy
  • Powered-By a Factory Replaceable 10-Year 3.6 Volt Lithium Battery
  • 25/second to 8 Hour User-Selectable Sampling Rate
  • Continuous (FIFO) or Stop When Full Sampling Methods
Requires ACR TrendReader 2 Software and USB Cable Kit

Due to Increased Freight Costs, All ACR Systems Orders Must Have a Minimum Order Value of $100

Special order - please allow 6-8 weeks for this item to ship.
User Manuals and Product Guides

SRP-001 SmartReader Plus 1 01-0008 Temperature Data Logger Specifications

Channels Two (One for the internal temperature sensor and one for an external temperature thermistor, switch status or resistance)
Memory Available in 32 KB (21,500 readings), 128 K (87,000 readings) or 1.5 MB (1,048,000 readings) versions
Size 107 x 74 x 22 mm. (4.2” x 2.9” x 0.9”)
Weight 110 grams (3.75 ounces)
Case Material Noryl Plastic
External Connector Removable screw-type terminal strip with Common (-) connection
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95%
Relative Humidity (non-condensing)
Clock Accuracy ±2 seconds per day
Battery 3.6 volt Lithium, 1 Amp-Hour
Battery Life 10 year warranty (under normal use), factory replaceable
Power Consumption 5 to 10 microamps (continuous)
Sampling Methods 1. Continuous (First-in, First-out – not available with
sample rates faster than 8 seconds)
2. Stop when full (Fill-then-stop)
Sampling Rates User selectable rates from 25 per second to 1 reading every 8 hrs. Battery Pack (#BP-101) or Power Supply (#PS-201) required for sample rates faster than 8 seconds.
Resolution 12 bit (1 part in 4,096)
Mounting Magnetic backing or locking eyelet
OS Compatibility Windows 11, 10 and 7 (32 and 64 Bit)
Software Requirements TrendReader 2 Software and Interface Kit
Product Approvals Meets Part 15 for Digital Devices of the Code of Federal
Regulations of the Federal Communications Commission
(FCC). Meets radiated and conducted emission require
ments for the Interference Causing Equipment Standard f
or Digital Apparatus of Industry Canada (ICES-003).
IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity,
EFT/Burst and Radiated Emissions respectively for the
Commission of the European Communities (CE)
Resistance to X-Rays Tested for protection against a 160kV dose @ 5mA for 30
seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times
that of an airport x-ray machine). Tested for protection
against Gamma Ray (equivalent to 0.137-1.38 mega volts)
IR 192 - 28 curies @ 30 second exposure source to object
distance 5”
Internal Temperature Sensor
Type Negative Temperature Coefficient Thermistor.
10,000 ohms @ 25°C (77°F)
Accuracy ± 0.2°C, over the range of 0 to 70°C (±0.3°F over the
range of 32 to 158°F)
Range -40 to 70°C (-40 to 158°F)
Resolution 0.03°C (0.05°F) at 25°C (77°F). Better than 0.07°C
(0.12°F) between -25°C (-13°F) and 70°C (158°F).
Better than 0.13°C (0.23°F) between -40°C (-40°F)
and -25°C (-13°F)
Customer Questions
Is the software compatible with Windows 10?
Yes, it will work with Windows10
Please, mind that only logged in users can submit questions
Copyright © 2024 MicroDAQ, LLC. All Rights Reserved.