  • ±0.5°C (±1.0°F) Accuracy
  • 3 Year Non-Replaceable Battery
  • Programmable Visual and Audible Alarms
  • Electronic Signature on Files to Meet FDA 21 CFR Part 11
  • Includes a 2-Year NIST Traceable Calibration Certificate, Compliant to ISO/IEC 17025:2017
  • Meets CDC & VFC (Vaccines for Children) Program Recommendations for Safe Vaccine Storage & Handling

Includes Attached USB Plug, CR-2032 Lithium Battery, 1.5M External Sensor with Glycol Buffer Vial and NIST Traceable Calibration Certificate.

Grouped product items
Description Inventory Price Qty
Fridge Version SKU: FRIDGE-TAG2L
In Stock
$176.46
− +
Freezer Version SKU: FRIDGE-TAG2L-FZR
In Stock
$176.46
− +
Ultra Low Version SKU: FRIDGE-TAG2L-ULTRA-LOW
In Stock
$249.00
− +
Backup Version SKU: FRIDGE-TAG2L-BKUP
In Stock
$186.00
− +
 

Overview of the Berlinger FridgeTag-2L Series Vaccine Monitors

The Berlinger Fridge-Tag2L Data Loggers are specifically built for monitoring stored vaccines in a refrigerators and freezers. The Fridge-Tag2Ls don’t require any software, all setup is done using two buttons on the face of the device. When connected to a computer using the integrated USB plug the data logger automatically creates a PDF report showing up to 56 days of recorded temperature readings. The easy to read, large, LCD displays current temperature, time, recording status, and battery life. Additionally, you can view up to 30 days of high, low, and any alarms and their duration just with the click of a button. In addition to the visual alarm indictor there is an audible alarm, ensuring that temperature excursions don’t go missed. The Fridge-Tag2L Vaccine Monitors use a flat ribbon sensor cable that minimizes air gaps and includes a highly accurate sensor mounted inside a 7ml buffer bottle filled with a Biosafe glycol solution.

The Fridge-Tag2L vaccine data loggers include the sensor with buffer bottle, mounting accessories, and a 2-year calibration certificate by our accredited ISO/IEC 17025:2017 in-house calibration lab.

The logger uses a non-replaceable battery that has a typical battery life of up to 3 years.

Berlinger Fridge Tag 2L Series Vaccine Data Logger Specifications

Temperature Measurement Range (Fridge, Freezer and Backup Versions): -40°C to 60°C (-40°F to 140°F)
Measurement Range (Ultra Low Version): -90°C to -5°C (-130°F to 23°F)
Accuracy (Fridge, Freezer and Backup Versions): ±0.5°C (-30°C to 40°C); ±0.8°C (Other Ranges)
Accuracy (Ultra Low Versions): ±1.5°C (±3.2°F)
NOTE: Display will only be visible in -10°C to 55°C (14°F to 131°F)
Sampling Rate 1 Minute
Logging Rate 5 Minutes
Alarms Fridge Version: Programable 1 High and 1 Low Alarm
Freezer Version: Fixed 1 High Alarm
Backup Version: Programable 1 High and 1 Low Alarm
Ultra Low Version: -70°C after 30 Minutes (High Alarm)
Temperature Range: -20°C to 50°C, in 0.1°C Increments
Alarm Time: 1 Minute to 1 Day
Type: Visual LED Indicator and Audible (Beep)
Typical Alarm Configuration Fridge Version: 60 Min Above 8°C (Upper Limit); 15 Minutes Below 2°C (Lower Limit)
Freezer Version:60 Minutes Above -15°C (Upper Limit); No Lower Limit
Data Storage 30 Day Overview on the Display
PDF Report Length: 56 Days
Date / Time Programming Data and Time Will Be Set When Device is Activated
Time Accuracy ±30 Minutes per Year
Activation Pressing the SET and READ Button for at Least 3 Seconds
Power Source CR-2032 Lithium Battery (Non-Replaceable) Complies with IATA DGR Packaging Instruction 970-Section 2 and is Not Considered a Dangerous Good
Battery Life Up to 3.5 Years (Typical Usage)
Calibration 2-Year ISO 17025:2017 Calibration Certificate (Included)
Fridge: 5°C
Freezer: -15°C
Communication Type USB Cable Connects Data Logger to PC or Mac
Time Accuracy ±30 Minutes/Year
Standards Compliance IP64, ASTM 2500, 21 CFR Part 11, 21 CFR Part 820, ICH Q8/Q9, ANSI/IEEE 829/830/1008, EN12830, 2002/95 EC Directive (RoHS), EN 61000-4-2/6-2/6-4 EMC, EN/IEC 60529 IP Protected Classes, IATA Regulations
Weight Data Logger: 3.95 oz. (112 grams)
Cable and Filled Vial: 0.77 oz. (22 grams)
Glycol Bottle Content of Biosafe Glycol: 7 ml
Operating Temperature: -40°C to 60°C (-40°F to 140°F)
LCD Display: 3.74" x 1.06" (9.5 cm x 2.7 cm)
External Cable Length: 1.5 Meters
Dimensions Data Logger: 5" x 3" x 0.75" (12.8 cm x 7.5 cm x 1.9 cm)
Vial: 1.02" x 1.02" x 1.33" (2.6 cm x 2.6 cm x 3.4 cm)
LCD Display: 3.74" x 1.06" (9.5 cm x 2.7 cm)
External Cable Length: 1.5 Meters