8 Channel Temperature Data Acquisition (DAQ) Module

Measurement Computing Module with Built-in Temperature Probe Accepting 8 Input Channels
SKU
USB-TEMP
In Stock
$629.00
  • (8) 24-Bit Analog Inputs for Thermocouple, RTD, Thermistor and Semiconductor Sensor Probes
  • (2) Integrated Cold Junction Compensation (CJC) Sensors
  • An Independent Alarm for Each Input
  • (8) User-Configurable Digital I/Os Configured for Input or Output

Includes USB Cable and Measurement Computing DAQ Software Suite (DAQami, TracerDAQ, InstaCal, Universal Libraries for Windows, Android and NI LabVIEW) Available as a Download

Software Downloads
zip
(Size: 32.96 MB)
exe
(Size: 361.69 MB)
Version 6.75
User Manuals and Product Guides
pdf
(Size: 0.01 MB)
This Document is Required to be Filled out by Anyone Outside the United States Purchasing a Measurement Computing Product.
Product Videos

Measurement Computing USB-TEMP USB Data Acquisition DAQ Module Specifications

Maximum Logging Rate/Channel 2 Samples per second
A/D Converters (4) Dual 24-bit Sigma-Delta type
Number of Channels 8 Differential Channels
Input Isolation 500 VDC minimum between field wiring and USB interface
Channel configuration Software programmable to match sensor type
Differential Input Voltage Range Thermocouple: ±0.080 Volts
RTD: 0 to 0.5 Volts
Thermistor: 0 to 2 Volts
Semiconductor Sensor: 0 to 2.5 Volts
Absolute Maximum Input Voltage Power On: ±25 Volt
Power Off: ±40 Volt
Input Impedance 5 Gigohm Minimum
Input Leakage Current Disabled Open Thermocouple Detect: 30 nA Maximum
Enabled Closed Thermocouple Detect:105 nA Maximum
Rejection Ratio Normal Mode: 90 dB Minimum
Common Mode: 100 dB Minimum
Resolution 24 Bits
Input Coupling DC
Warm-up Time 30 Minutes Minimum
Open Thermocouple Detect Automatically enabled when the channel pair is configured for thermocouple sensors. The maximum open detection time is 3 seconds.
CJC Sensor Accuracy 15°C to 35°C: ±0.25 °C (Typical); ±0.5 °C (Maximum)
0°C to 70°C: -1.0 to +0.5 °C (Maximum)
Digital Input/Output Type CMOS
Number of Digital I/O 8 (DIO0 to DIO7)
Digital I/O Configuration Independently configured for input or output.
Power on reset is input mode.
Digital Pull up/Pull down Configuration All pins pulled up to +5 V via 47 K resistors (default). Pull down to ground (GND) also available.
Digital I/O Transfer Rate
(software paced)
Digital Input: 50 port reads or single bit reads per second typical.
Digital Output: 100 port writes or single bit writes per second typical.
Digital Input High Voltage: 2.0 V (Minimum); 5.5 V (Absolute Maximum)
Low Voltage: 0.8 V (Maximum); -0.5 V (Absolute Minimum)
Digital Output Low Voltage (IOL = 2.5 mA): 0.7 V (Maximum)
High Voltage (IOH = -2.5mA): 3.8 V (Minimum)
Memory Storage Format PC Memory
Power Self-Powered USB Port
Power Supply Current USB Enumeration: <100 mA
Continuous Mode: 70 mA Maximum
USB Cable Type A-B cable, UL type AWM 2527 or equivalent (min 24 AWG VBUS/GND, min 28 AWG D+/D-)
USB Cable Length 3 meters Maximum
Operating Range Temperature: 0°C to 70°C
Humidity: 0 to 90% non-condensing
Storage Range Temperature: -40°C to 87°C
Humidity: 0 to 90% non-condensing
OS Compatibility Windows 11, 10 and 7 (32 and 64-Bit) for Universal Library, LabVIEW
Dimensions 127mm x 88.9mm x 35.56mm
Customer Questions
what standard need us to follow for yearly calibration? Thanks
Calibration can be purchased as a separate option when adding the logger to the shopping cart.  We default to a 1 year calibration interval unless otherwise specified by the customer.  We can recalibrate the logger for you once the certificate...
Please, mind that only logged in users can submit questions