MSR145 Miniature Signal Data Recorder

MSR Electronics Data Recorder Measures and Records Temperature, Humidity, Pressure, Light, Acceleration and 2 or 4 Additional Analog Inputs Available
SKU
MSR145
Please open the Configurator tab to choose your desired options.
  • Measures Temperature, Humidity, Light, Pressure and Acceleration Simultaneously and 2 or 4 Analog Inputs Also Available
  • Stores up to 2 Million (Internal Memory) or 1 Billion (SD Card) Total Readings
  • Rechargeable 230 or 1000 mAh Lithium Polymer Battery
  • Data from any number of loggers can be merged synchronously into a single Data Record
  • Accelerometer Range of ±15g with -20°C to 65°C Temperature Range
  • Made in Switzerland, Swiss Made  Swiss Made - Made in Switzerland

Includes Software CD, USB Cable and 4GB microSD Card

Click on the "Configurator" Tab Below to Choose the Ordering Options

MSR Data Loggers are Built to Order and Not Eligible for Return. All Sales Are Final.
Special order - please allow 4 weeks for this item to ship.

With a few clicks you are able to select your desired MSR145.

  1. Select the housing and the corresponding battery.
  2. Select the configuration.
  3. Would you like to expand the memory capacity to over 1 billion measured values?
  4. General technical data of the MSR145 data logger

1. Select the housing and the corresponding battery.

Housing Type and battery Picture Size & weight Medium Sheat
material
Item no.
Standard
IP 60
Design housing
with 230 mAh battery
MSR145S 27 x 16 x 53 mm
(W x H x L)
approx. 20 g
Air PC, not
encapsulated
145B15
Error
Design housing
with 1000 mAh battery
MSR145 900 mAh 39 x 23 x 72 mm
(W x H x L)
approx. 56 g
Air PC, not
encapsulated
145B7
Error
Waterproof
IP 67
Silicon tube
with 230 mAh battery
MSR145W 20 x 14 x 62 mm
(W x H x L)
approx. 18 g
Air,
various liquids
Silicon,
encapsulated
145B4
Error
Design housing
with 230 mAh battery
MSR145WS 27 x 16 x 53 mm
(W x H x L)
approx. 27 g
Air,
various liquids
PC,
encapsulated
145B16
Error
Design housing
with 1000 mAh battery
MSR145 900 mAh 39 x 23 x 72 mm
(W x H x L)
approx. 64 g
Air,
various liquids
PC,
encapsulated
145B8
Error

2. Select the configuration.

You can select various internal sensors as well as up to five external sensors or connectors.

Measured parameter /
Measurement/
storage rate
Technical data Internal External (according to cable length)
0.20 m 1.00 m 1.60 m
Temperature

1/s to every 12 h
Working range:
external: -40...+125 °C
internal: -20...+65 °C

Max. deviation:
±0.1 °C (-25...+55 °C)
±0.3 °C (-40...+125 °C)
Item no. T9
Error
Item no. FT9/020
Error

Item no. FT9/100
Error

Item no. FT9/160
Error

Working range:
-55...+125 °C

Max. deviation:
±0.5 °C (10...+70 °C)
±2 °C (-55...+125 °C)
only available
externally
Item no. FT2/020
Error

Item no. FT2/100
Error

Item no. FT2/160
Error

Connector for thermal
element of Type K

(without sensor)


4/s* to every 12 h
*(1/s in combination
with A4 sensor)
Working range:
-200...+1350 °C
only available
externally
Item no. FT3/020
Error

Item no. FT3/100
Error

Item no. FT3/160
Error

Humidity
with integrated
temperature

1/s to every 12 h
Working range:
0...100 % rel. humidity
external: -40...+125 °C
internal: -20...+65 °C

Max. deviation:
±2 % rel. humidity
(10...80 %, +25 °C)
±3 % rel. humidity
(80...100 %, +25 °C)
Item no. H3
Error
Item no. FH3/020
Error

Item no. FH3/100
Error

Item no. FH3/160
Error

Air pressure absolute,
with integrated
temperature

Internal:
1/s to every 12 h
External:
10/s* to every 12 h
*(1/s in combination
with A4 sensor)
Working range:
10...2000 mbar absolute
external: -40...+85 °C
internal: -20...+65 °C

Max. deviation:
±2 mbar (1% FS)
(750...1100 mbar absolute,
+25 °C)
Item no. P5
Error
Item no. FP10/020
Error

Item no. FP10/100
Error

Item no. FP10/160
Error

Working range:
0...14 bar absolute
-20...+65 °C

Max. deviation:
±50 mbar (1% FS)
(1...10 bar abs., +25 °C)
Item no. P7

only available
on request
Item no. FP7/020
Error

Item no. FP7/100
Error

Item no. FP7/160
Error

Fluid pressure
with integrated
temperature

Internal:
1/s to every 12 h
External:
10/s* to every 12 h
*(1/s in combination
with A4 sensor)

Material in contact
with media:
  • Stainless Steel AISI 316L
    (DIN 1.4404 / 1.4435)
  • O-rings: Viton® 70° Sh
Working range:
-1000...3000 mbar relative
-20...+85 °C

Max. deviation:
±30 mbar (1% FS)
only available
externally
Item no. FP9/020
Error

Item no. FP9/100
Error

Item no. FP9/160
Error

Working range:
0...3000 mbar absolute
-20...+85 °C

Max. deviation:
±30 mbar (1% FS)
only available
externally
Item no. FP3/020
Error

Item no. FP3/100
Error

Item no. FP3/160
Error

Working range:
0...30 bar absolute
-20...+85 °C

Max. deviation:
±0.3 bar (1% FS)
only available
externally
Item no. FP4/020
Error

Item no. FP4/100
Error

Item no. FP4/160
Error

Working range:
0...400 bar absolute
-20...+85 °C

Max. deviation:
±4 bar (1% FS)
only available
externally
Item no. FP8/020
Error

Item no. FP8/100
Error

Item no. FP8/160
Error

Working range:
0...1000 bar absolute
-20...+85 °C

Max. deviation:
±10 bar (1% FS)
only available
externally
Item no. FP12/020
Error

Item no. FP12/100
Error

Item no. FP12/160
Error

3-axis-acceleration
(static)

50/s to every 12 h
±15 g
-20...+65 °C

±0.15 g (0...5 g, +25 °C)
±0.25 g (5...10 g, +25 °C)
±0.45 g (10...15 g, +25 °C)
Item no. A
Error
Item no. FA/020
Error
Item no. FA/100
Error
Item no. FA/160
Error
3-axis-acceleration
with Fast Peak (1 kHz)


50/s to every 12 h
Item no. A4
Error
Item no. FA4/020
Error
Item no. FA4/100
Error
Item no. FA4/160
Error
Light

1/s to every 12 h
Working range:
0...83 000 lx

Max. sensitivity
at 550 nm
Item no. L3
Error
Item no. FL3/020
Error

Item no. FL3/100
Error

Item no. FL3/160
Error

Connection for
strain gauges
Characteristics:
  • full bridge
  • rated output 1 mV/V
  • bridge voltage 3 V
only available
externally
Item no. FD/020
Error

Item no. FD/100
Error

Item no. FD/160
Error

Analogue inputs
(resolution: 12 bit)

50/s* to every 12 h
*(1/s in combination
with A4 sensor)
Input configuration:




only available
externally
Exxxx/020
Error
Exxxx/100
Error
Exxxx/160
Error
Control terminal Characteristics:
  • Alarm output
  • Input for starting and
    stopping data recording
  • 3 V voltage output
only available
externally
Item no. FY/020
Error
Item no. FY/100
Error
Item no. FY/160
Error

3. Would you like to expand the memory capacity to over 1 billion measured values?

Please order the expansion slot for plugging-in a standard microSD card here. Item no.
Expansion slot for microSD card (available only for design housing B7 and B8)

Please note that:
  • this expansion slot with plugged-in microSD card only meets the protection category IP60,
    even with the design housing B8,
  • one formatted and ready for use microSD card (≥4GB) is included.
N10006
Error

Your selection:

Price:
Item no.:

MSR145 Multifunctional Data Recorder Specifications

Measured Parameters Temperature
Relative Humidity (RH) with Integrated Temp
Air Pressure
Fluid Pressure
3-Axis Acceleration (ie. Determining Position)
Light
2 or 4 Analog Inputs (0 to 3.0 Volts, 12-bit) - Standard
Note: Additional Ranges Available

0-20mA, 4-20mA, 0-3, 5, 10, 12, 24, 36, or 48 volt

Acceleration Range: ±15g @ -20°C to 65°C
Accuracy: ±0.15G @ 25°C
Resolution: 4mg
Response Time: 4mg
Temperature Range (Internal): -20°C to 65°C
Range (External): -10°C to 58°C; -55°C to 125°C; 0 to 1024°C (External Type K TC Connectors)
Accuracy (Internal): ±0.5°C
Accuracy (External): ±0.1°C (5°C to 45°C); ±0.2°C (-10°C to 58°C); ±0.5°C (-10°C to 65°C), ±2.0°C (-55°C to 125°C)
Resolution: 0.034°C
Response Time: Approximately 30 Seconds
Humidity Range: 0 to 100% RH, -20°C to 65°C
Accuracy: ±2% (10 to 85% RH, 0°C to 40°C); ±4% (85 to 95% RH, 0°C to 40°C)
Air Pressure Range: 0 to 2000 mbar Absolute or 0 to 14 Bar Absolute
Accuracy: ±2.5mbar (750 to 1100 mbar Absolute, 25°C); ±50 mbar (1-10 bar Absolute, 25°C)
Resolution: 15 Bit
Response Time: Approximately 1/10 Second
Fluid Pressure Range: 0 to 3000 mbar Absolute or 0 to 30 Bar Absolute
Accuracy: ±30mbar; ±300 mbar
Resolution: 15 Bit
Response Time: Approximately 1/10 Second
Light Range: 0 to 65,000 lx
Accuracy: 100 to 65,000 lx, Max ±5% Linearity Error, Max Sensitivity at 500 nm
Resolution: 4 lx
Analog Input Number: 2 or 4 (Each Channel Configured for Separate Ranges)
Range: 0 to 3.0 Volts (Standard)**
Resolution: 12 Bits
Note: Additional Ranges Available
0-20mA, 4-20mA, 0-3, 5, 10, 12, 24, 36, or 48 volt
Sampling Rate Temperature & Humidity: 1/s to 12 Hours
Pressure: 10/s to every 12 Hours
Acceleration and Analog Input: 50/s to 12 Hours
Light Level: 1/s to 12 Hours
Analog: 50/s to 12 Hours
Data Storage Capacity Internal Storage: Approximately 2 Million Total Readings
External SD Card: Approximately 1 Billion Total Readings
Operating Mode Push Button Start/Stop
Storage Conditions Temperature: 5°C to 45°C
Humidity: 10 to 95% RH, non-condensing
LEDs Blue: Recording Indicator
Red: Alarm Indicator
Yellow: Battery Charge Indicator
Power Supply Rechargeable Lithium Polymer Battery (260mAh or 900mAh)
(USB Connection to PC recharges unit)
Battery Life 2-3 Months (Sampling Rate: 1 Minute)
OS Compatibility Windows 11, 10 and 7 (32 and 64 Bit)
Enclosure Standard Version: Poly Carbonate, Non-Molded
Waterproof Version: Silicon, Molded
Standard Waterproof Version: Poly Carbonate, Molded
Media Standard Version: Air
Waterproof Version: Air, Various Liquids
Standard Waterproof Version: Air, Water
Standards Compliance RoHS/WEEE, IP60 (Standard Version), IP67 (Waterproof Versions)
Dimensions Standard Version: 20mm x 15mm x 52mm
Waterproof Version (Silicon Tube): 18mm x 14mm x 62mm
Standard Waterproof Version: 20mm x 15mm x 61mm
Weight Standard Version: 16 grams
Waterproof Version: 18 grams
Standard Waterproof Version: 23 grams
Software Downloads
zip
(Size: 255.38 MB)
Version 6.09.01 Includes ShockViewer Add-On for MSR165 and MSR175.
User Manuals and Product Guides
pdf
(Size: 3.05 MB)

Mounting Instructions for MSR Shock and Vibration Data Loggers

The data loggers should be force-fitted as close as possible to or on the object to be monitored. Fastening to the packaging or the means of transport will lead to deviating results.

Screw the data loggers to the object at the holes provided. Alternatively, the data loggers can also be fastened with high-strength industrial adhesive tape or cable ties. Magnetic fastening should be avoided. It only acts in the direction of the magnetic field lines - it is not force-locking.

Mounting Instructions for MSR Shock and Vibration Data Loggers
User Manuals and Product Guides

Orientation

Align the Axes of the data Loggers with the main direction of movement. The data loggers can be mounted in any position. Make a note of the position and direction of the axes for later evaluation.

MSR Mounting
Customer Questions
Is the software compatible with Windows 10?
Yes, it will work with Windows10
Can one of the analog inputs be used as a trigger to capture acceleration events? For example, record 5 seconds before the trigger (falling and/or rising voltage) to 5 seconds after the trigger event?
Yes, this meets the requirements of the VFC.  All of our vaccine monitoring loggers meet their requirements.  This particular logger has audible alarm.  