MSR145 Bluetooth Wireless Data Logger Specifications
Measured Parameters
Temperature Relative Humidity (RH) with Integrated Temperature Air Pressure with Integrated Temperature and Acceleration (ie. Determining Position) Light
Temperature
Range: -20°C to 65°C Accuracy: ±0.2°C (-10°C to 50°C); ±1.0°C (-40°C to 125°C) Note: Available in C Only.
Humidity
Range: 0 to 100% RH @ -20°C to 65°C Accuracy: ±1.8% (10 to 85% RH); ±4% (85 to 95% RH)
Air Pressure
Range: 0 to 2000 mbar Absolute @ -20°C to 65°C Accuracy: ±2 mbar (750 to 1100 mbar Absolute, 25°C)
Acceleration
Range: ±15g @ -20°C to 65°C Accuracy: ±0.15G @ 25°C
Light
Range: 0 to 65,000 lx Accuracy: 100 to 65,000 lx, Max ±5% Linearity Error, Max Sensitivity at 500 nm
Bluetooth
Type: Bluetooth 4.0 Smart Frequency: 2.4 GHz Radio Range: Approximately 10 Meters Note: Communication with Multiple Data Loggers at the Same Time
Sampling Rate
1 Second to 12 Hours
Data Storage Capacity
Approximately 1 Million Total Readings
Display
Type: High Contrast OLED (Organic Light-Emitting Diode) Resolution: 96 x 64 pixels Display Area: 22.14 mm x 15.42mm Fast Response Time: <10 µs Wide Viewing Angle:>160°
Mounting Instructions for MSR Shock and Vibration Data Loggers
The data loggers should be force-fitted as close as possible to or on the object to be monitored. Fastening to the packaging or the means of transport will lead to deviating results.
Screw the data loggers to the object at the holes provided. Alternatively, the data loggers can also be fastened with high-strength industrial adhesive tape or cable ties. Magnetic fastening should be avoided. It only acts in the direction of the magnetic field lines - it is not force-locking.
Mounting Instructions for MSR Shock and Vibration Data Loggers
Align the Axes of the data Loggers with the main direction of movement. The data loggers can be mounted in any position. Make a note of the position and direction of the axes for later evaluation.