Conductivity Temperature and Water Level Depth Sensor up to 251ft

SKU
W-CTD-03
$895.00
  • Interchangeable Sensor Measures Conductivity, Temperature and Water Level
  • 0 to 100,000 µS/cm, -20C to 50C and 0 to 123 psia Measurement Ranges
  • 4-Electrode Conductivity Sensor for Wide Measurement Range
  • Calculates Conductance, Sallinity and TDS (Total Dissolved Solids)
  • Accurate Shallow Water Levels with Durable Ceramic Sensor in Tip
  • Depth Level Range: 0 to 51 ft
Special order - please be advised that due to supply shortages, some items may require 6 weeks for shipment

Onset HOBO W-CTD Conductivity Temperature and Depth Water Sensor Specifications

Conductivity Measurement Range (Electrical Conductivity and Specific Conductance): 0 to 100,000 µS/cm
Measurement Range (Salinity using PSS-78): 2 to 42 PSU
Measurement Range (Total Dissolved Solids): 0 to 100,000 mg/L
Calibrated Range (Electrical Conductivity): 50 to 80,000 µS/cm @ 5°C to 35°C (41°F to 95°F)
Accuracy (Conductivity): ±2% or 15 µS/cm Whichever is Greater
Accuracy (Salinity): ±2% of Reading or 0.1 PSU, Whichever is Greater
Resolution (Conductivity): 0.1 µS/cm from 0 to 1,000 µS/cm; 1 µS/cm from 1,000 to 10,000 µS/cm; 10 µS/cm from 10,000 to 100,000 µS/cm
Resolution (Salinity): 0.01 PSU
Resolution (Total Dissolved Solids): 0.1 mg/L
Response Time: 1 second to 90% at a Stable Temperature
Absolute Pressure Measurement Range: 0 to 850 kPa (0 to 123.3 psia)
Factory Calibrated Range: 69 to 850 kPa (10 to 123.3 psia) @ 0°C to 40°C (32°F to 104°F)
Burst Pressure: 1200 kPa (174 psia) or 112 Meters (368 ft) Depth
Raw Pressure Accuracy*: Maximum Error - ±0.3% FS, 2.55 kPa (0.37 psi)
Pressure Response Time**: <1 Second to 90% at a Stable Temperature
* Raw Pressure Accuracy: Absolute Pressure Sensor Accuracy includes All Sensor Drift, Temperature and Hysteresis-Induced Errors.
** Changes in Temperature: Allow 20 Minutes in Water to Achieve Full Temperature Compensation of the Pressure Sensor. There can be up to 0.5% of Additional Error due to Rapid Temperature Changes.
Water Level Measurement Range: 0 to 76.5 Meters (0 to 251 ft) of Water Depth at Sea Level or 0 to 79.5 Meters (0 to 262 ft) of Water at 3,000 Meters (10,000 ft) of Altitude
Accuracy*: Typical - ±0.05% FS, 3.8 cm (0.125 ft) Water; Maximum - ±0.1% FS, 7.6 cm (0.25 ft) Water
Resolution: <0.085 kPa (0.012 psi), 0.87 cm (0.028 ft) Water
* Water Level Accuracy: With Accurate Reference Water Level Measurement, Known Water Density, and a Stable Temperature Environment. System Water Level Accuracy Equals the Sum of the Barometric Water Level Accuracy Plus the Selected CTD Sensor Water Level Accuracy.
Temperature Measurement Range: -20°C to 50°C (-40°F to 122°F), Non-Freezing Water
Accuracy: ±0.15°C (±0.27°F) from 0°C to 50°C (32°F to 122°F)
Resolution: 0.01°C at 25°C (0.018°F at 77°F)
Response Time: 3 minutes to 90% in Water (Typical)
Drift: <0.1°C (0.18°F) per Year
Wetted Materials Passivated 316 Stainless Steel Housing Rated for Use in Saltwater, Viton and Buna-N O-Rings, PET Sensor Connector
Depth Sensor: Ceramic
Conductivity Sensor: PET, Platinum Plating on Electrodes
Note: Sensor Should be Mounted so That it is Not in Contact with Other Metals
Standards Compliance FCC (US), CE (EU), UKCA, Waterproof to Pressure Sensor Burst Pressure Rating, IP68