4-Channel Thermocouple Data Logger w/ Graphing LCD Display
- Measure and Record Temperature Using Type K, J, T, S or R Thermocouples (Sold Separately)
- Data Transfered to PC for Analysis Using USB (Cable Included) or SD Card
- Touch Panel for Setup and Configuration Changes in the Field
- LCD Display Shows Current Readings, Trend Graph, Battery Status and Configuration Settings
- Stores up to 240,000 Temperature Readings per Channel (960,000 Total)
Includes Batteries, USB Cable, PC Software, Card Slot Cover and User Manual
TandD MCR-4TC 4-Channel Thermocouple Data Logger
The TandD MCR-4TC is a 4-channel thermocouple data logger that is easy to use and very reliable. This battery-operated four channel multi-channel data logger can be used to measure temperatures in boilers, cooking equipment (ovens, fryers), ducts and pipes, in addition to managing frozen or refrigerated foods and temperature testing electronic substrates. It captures a wide temperature range (-454°F to 3200°F (or -270°C to 1760°C) with a sample rate that is selectable from 10Hz to 1 hour. Data is recorded to an SD card and can be transferred to a computer via the SD card or a USB cable. The logger features an easy to read LCD display which has real-time graphing capability. Up to four loggers can be docked together to create a 16 channel logger that can be controlled from a single unit.
4-Channel Thermocouple Data Logger Specifications
| Thermocouple | Number of Channels: 4 External Inputs for Thermocouples Type: J, K, T, R and S Connection Type: Screwless Terminals Cold Junction Compensation: Automatic Measurement Resolution: 0.1°C Input Impedance: 1MΩ (Approximate) |
| Thermocouple Types | Type J: -210°C to 1200°C @ ±0.5°C + 0.3% (Accuracy) Type K: -270°C to 1370°C @ ±0.5°C + 0.3% (Accuracy) Type T: -270°C to 400°C @ ±0.5°C + 0.3% (Accuracy) Type R: -50°C to 1760°C @ ±1.5°C + 0.3% (Accuracy) Type S: -50°C to 1760°C @ ±1.5°C + 0.3% (Accuracy) |
| Cold Junction Compensation | Accuracy: ±0.5°C (10°C to 40°C) ; ±0.8°C (Other Ranges) |
| Recording Method | Instanteous, Average (Ave), Fine |
| Sampling Interval | 100, 200, 500 ms 1, 2, 5, 10, 15, 20, 30 Seconds 1, 2, 5, 10, 15, 20, 30, 60 Minutes |
| Data Storage Capacity | Up to 960,000 Readings (1 Channel); Up to 480,000 Readings (2 Channels); Up to 320,000 (3 Channels); Up to 240,000 Readings (4 Channels) |
| Recording Mode | Endless: Overwrite Oldest Data when Capacity is Full One-time: Stop Recording when Capacity is Full |
| Group Recording | Up to 4 Units (16 Channels) Can Be Recorded Simultaneously |
| Start Modes | Immediate or Programmed |
| LCD Display | Current Readings, Battery Life and Trend Graph |
| Power | (2) AA (Alkaline or Lithium) Battery USB Powered |
| Battery Life | At 100ms Sample Rate : 5 Day Typical At 500ms Sample Rate : 7 Day Typical At 1 Second Sample Rate : 7 to 21 Day Typical At 5 Second Sample Rate or Longer : 21 to 60 Day Typical NOTE: Battery Life Depends on Measurement Environment, Sampling Interval and Battery Performance |
| Communication Type | USB |
| External Storage | SD Memory Card (SDHC Memory Card) - Sold Separately NOTE: Measurement Data Cannot be Directly Recorded onto the Memory Card. Used to Transfer Recorded Data from Data Logger to PC Only. |
| Input Terminals | Screwless Terminals Single Wire: 0.32mm to 0.65mm (28 AWG to 22 AWG) Twisted Wire: 0.08 to 0.32mm (28 to 22 AWG). 0.12mm or More in Diameter Strip Length: 9mm to 10mm |
| Isolation | CH1, CH2, CH3, CH4, USB are Isolated Max Applied Voltage: ±50 Volts Electrical Isolation Resistance: 50 MΩ or More NOTE: Battery Terminals are Not Isolated from CH1 to Ch4 Input Terminals |
| Operating Environment | Temperature: 0°C to 50°C Humidity: 90%RH or Less (No Condensation) |
| Weight | 190 grams (Includes Battery) |
| Dimension | 120mm x 75mm x 32mm |