LogTag CP100 Calibratable Smart Probe Kit
- -40°C to 90°C Measurement Range with 0.8°C Accuracy
- Sensor Probe Can Be Swapped Out and Calibrated Separately
- Gold-Plated MCX Connector for Secure Connection
- 3 Year Battery Life
- Compatible with the LogTag UTRED30-16 and UTRED30-WIFI Data Loggers
Kit Includes CP100K-15 Sensor, 60ml Glycol Bottle with Base. ISO Calibrated for Fridge and Freezer
The CP100K-15-KIT is a precise, swapable, temperature probe that comes pre-calibrated for use in both fridges or freezers. Because the sensor itself is taking the temperature reading it eliminates the need to send in the actual logger when calibration becomes due. When calibration is due, simply hot swap the existing standard or smart sensor with a newly calibrated smart sensor ensuring zero downtime. These sensor kits include the sensor, glycol buffer bottle, mounting stand, and a calibration certificate. The sensors are sold as stand-alone items so that they can be purchased as additional sensors to use to replace existing expired ones. Additionally, these sensors can be purchased in various vaccine monitoring kits found elsewhere on our website.
The CP100K-15-KIT is compatible with the following loggers:
VAC-UTRED30-16 Logger & Kits
VAC-UTRED30-WIFI Loggers & Kits
LogTag CP100K-15-KIT Calibratable Smart Sensor Kit Specifications
Temperature | Measurement Range: -40°C to 90°C (-40°F to 210°F) Accuracy (-40°C to -20°C): ±0.5°C (±0.9°F) Accuracy (-20°C to 40°C): ±0.8°C (±1.4°F) Accuracy (40°C to 70°C): ±0.8°C (±1.4°F) Accuracy (70°C to 99°C): ±1.2°C (±2.2°F) Resolution (-40°C to 50°C): 0.1°C (0.2°F) Resolution (50°C to 70°C): 0.2°C (0.4°F) Resolution (70°C to 80°C): 0.3°C (0.5°F) Resolution (80°C to 99°C): 0.4°C to 0.6C (0.7°F to 1.0°F) |
Acquisition Interval | Determined by Logger, Shortest Acquistion Interval 30 Seconds |
Sensor Reaction Time | Based on Tip, Same as for Current Probes |
Operating Environment | Temperature (Probe End): -40°C to 99°C (-40°F to 210°F) Temperature (Signal Conditioning End): -20°C to 70°C (-4°F to 158°F) Humidity: 0 to 100% RH (Non-Condensing) |
Storage Environment | Temperature: -40°C to 99°C (-40°F to 210°F) Humidity: 0 to 100% RH (Non-Condensing) |
Connection Interface | 1-Wire Digital Interface to Compatible Data Loggers |
Construction | Signal Conditioning Case Material: Polycarbonate Tip Material: Stainless Steel 316 Cable Material: PTFE |
Environmental | Probe Tip: IP68 Probe Tip/Cable Junction: IP64 Signal Conditioning: IP44 |
Compatibility | UTRED30-WIFI Dual-Channel UTRED30-16 Dual-Channel UTRED30-16 Single-Channel |
Weight | 34 grams (1.2 oz.) |