VAC-UTRED30 Vaccine Monitoring Kits
- Measures and Records Temperature Readings Ranging from -40°C to 99°C Using External Sensor(s)
- Alarms can be Programed Individually for Fridge and Freezer allowing a single logger to monitor both a fridge and freezer simultaneously
- Powered by AC-Powered USB Plug and User-Replaceable AAA Alkaline Batteries with 1 Year Battery Life
- Stores Up To 32,000 Recorded Samples/Readings
- Built-in Audible Alarm and LCD Display with 30 Day Summary
- Each Sensor Alarm Can Be Configured Separately for Fridge and/or Freezer
- Includes a 2-Year NIST Traceable Calibration Certificate, Compliant to ISO/IEC 17025:2017
- Data Loggers are Calibrated for Both Fridge (5°C) and Freezer (-15°C) Use
- Kits are Available with a Standard ST100K Probe or Hot Swappable LogTag CP100K Smart Probes that can Calibrated on their own when Replaced
Includes UTRED30-16 Data Logger, sensor(s) with 60ml Glycol Bottle, Base for Bottle, Wall Mount Bracket , AC Adapter, (2) AAA Alkaline Batteries, Logger Installation Kit and NIST Traceable Calibration Certificate for Fridge and Freezer.
Overview of the UTRED30 Vaccine Monitoring Kits
The LogTag VAC-UTRED30 Vaccine Monitoring Kits are a complete solution for monitoring vaccines stored in both refrigerators and freezers. The large easy to read display is designed to show ‘at a glance’ if temperature excursions have occurred during the current day and up to the previous 29 days. Alarms are customizable during configuration, which is handled by the powerful, easy to use, free LogTag Analyzer software. The logger can be AC powered using the included AC adaptor and USB cable as well as by 2 AAA batteries. The logger also features a backup coin cell battery.
The VAC-UTRED30 Kits are available with a standard ST100K Probe or Hot Swappable Pre-Calibrated CP100 Sensors with 1.5 or 3.0 Meter Cable Lengths. At the heart of the kits are the LogTag UTRED30-16 data logger which has inputs for two external probes allowing you to monitor both a fridge and freezer with a single device. Additional probes can be purchased separately.
The kits includes everything you need to meet state health departments and the CDC for safe vaccine monitoring and storage. It includes the logger, sensor probe with glycol buffer bottles, calibration certificate, and mounting accessories.
LogTag VAC-UTRED30 Vaccine Monitoring Kits Specifications
| Temperature | Sensor Type: Precision Thermistors Sensors: ST100K-15, CP100K-15 or CP100K-30 Measurement Range: -40°C to 99°C (-40°F to 210°F) Accuracy (-20°C to 40°C): ±0.3°C (Typically) Accuracy (-40°C to -20°C): ±0.5°C (Typically) Accuracy (40°C to 70°C): ±0.7°C (Typically) Accuracy (70°C to 99°C): ±1.0°C (Typically) Resolution: 0.1°C or 0.1°F Reaction Time: Typically less than 2 Minutes (T90) in Moving Air (1 m/s) |
| Data Storage Capacity | 32,000 Total Recorded Readings/Samples (16,000 Per Sensor/Channel) |
| Sampling Frequency | 30 Seconds to 18 Hours |
| Alarms | (6) Individually Configurable Alarms for Each Sensor/Channel |
| Operating Mode | Push Button Start Specific Date & Time Optional Start Delay up to 72 Hours |
| Recording Indicators | Flashing "OK" Indicator Flashing "Alert" Indicator |
| External Probes | Standard ST100K and Hot Swappable Self-Calibrating CP100K Smart Probes Available Standard ST100K: Available in 1.5 Meter (#ST100K-15) or 3.0 Meter (#ST100K-30) Cable Length Self-Calibrating CP100K: Available in 1.5 Meter (#CP100K-15) or 3.0 Meter (#CP100K-30) Cable Length |
| Download Time | Typically with full memory (32,000 readings) in less than 20 secs depending on computer or readout device used |
| Standards Compliance | IP61, CFR21 Part 11, RoHS |
| Real Time Clock | Rated Accuracy: ±25ppm @ 25°C (Equivalent to 2.5 seconds/day) Rated Temperature Coefficient: -0.034 ±0.006ppm/°C (Typically ±0.00294 Seconds/Day/°C) |
| Power Source | AC Powered: 5 Volt Power Input from Micro-USB Cable Battery-Powered: (2) AAA User-Replaceable Alkaline Battery |
| Battery Life | Typically 18 Months of Normal Use @ 6 Minute Sampling Interval and Monthly Data Downloads |
| Operating Range | Temperature: -30°C to 60°C (-22°F to 140°F) |
| Software Compatibility | Windows 11, 10 and 7 (32 and 64-Bit) |
| Case Material | Polycarbonate |
| EMC Compliance | Tested and complies with EC EMC directives (EN 50081-1:1992 & EN 61000-6-1:2001) Tested and complies FCC Part 15 Subparts A and B. |
| Weight | 156g with AAA Batteries (without External Probes) |
| Dimensions | 141mm x 72mm x 18mm |